14 Agilent DNA 7500 and DNA 12000
5 Agilent DNA 7500 and DNA 12000 Assay Protocol
Loading the Marker
5 Press the plunger of the syringe down until it is
held by the clip.
6 Wait for exactly 30 seconds and then release the
plunger with the clip release mechanism.
7 Visually inspect that the plunger moves back at
least to the 0.3 ml mark.
8 Wait for 5 seconds, then slowly pull back the
plunger to the 1 ml position.
9 Open the chip priming station.
10 Pipette 9.0 µl of the gel- dye mix in each of the
wells marked
Loading the Marker
1 Pipette 5 µl of green- capped DNA marker vial
(green ) into the well marked with the ladder
symbol and into each of the 12 sample wells.
pressurize
9 µl gel-dye
NOTE
Protect the gel-dye mix from light. Store the gel-dye mix at 4 °C when not in use for more
than 1 hour.
5 µl marker
NOTE
Do not leave any wells empty or the chip will not run properly. Add 5 µl of green-capped
DNA marker (green ) plus 1 µl of deionized water to each unused sample well.